Issue StoriesProcedural Considerations for a Loudness Perception Profile for Childrenby Andrea Bohnert, MTA, & Petra Brantzen A look at the Phonak Loudness Perception Profile (LPP) fitting tool and its application in pediatric hearing instrument fittings. Over the last several years, measures of loudness growth have experienced increased research interest.1-3 This is primarily due to the increased use of non-linear amplification, the availability of more sophisticated signal processing strategies, and the availability of instruments that are highly flexible,4 thus allowing the hearing care professional to better meet individual amplification needs. Advances in digital signal processing have led to the widespread use of instruments with sophisticated signal processing technology. These devices have proven to be highly successful in the adult population. Increasingly, they are also recommended for use with children. However, data regarding the performance of children with these products is still scarce. Modern digital hearing systems require the configuration of a larger number of parameters than their analog predecessors. To take advantage of this flexibility, it is important to select and pre-set amplification based on the full range of unique characteristics and abilities of each child.5 The goal of a pediatric fitting is to provide access to the widest possible range of auditory information. In order to ensure this, any additional tools that help the clinician gain maximum audiological information while maintaining the childs interest and cooperation is desirable. Interactive fitting procedures, such as loudness scaling, have proven to be useful in presetting hearing instruments more accurately to individual needs.6 They not only provide a better understanding of individual, ear-specific and frequency-specific dynamic range, but also avoid the inaccuracies of measuring uncomfortable loudness levels.7 One such fitting tool is the Loudness Perception Profile (LPP) developed by Phonak. This software-based procedure has been used in adults for some time.6 Anecdotal evidence exists about the success of this procedure in children,8 but no clinical data has been collected to fully evaluate the use of this tool for fitting children. The following field study was designed to examine whether the procedure can be reliably applied to children ages seven years and older.
Subjects Procedure Phonak Claro 211 and 311 dAZ hearing instruments were used in the trial to generate the signal. For the hearing-impaired group, the childs own custom earmolds were used, and soft earmolds with 0.8 mm vents were made for the control group of children with normal hearing.
Responses were automatically input to the fitting software by means of the Active Response Unit (ARU). The ARU displays a 10+1 point scale from inaudible (+1) to extremely loud (10), and subjects are asked to judge the loudness of the presented stimuli. They indicate their choices for selected loudness categories directly on the response unit. A special pediatric scale that uses pictures (Fig. 2) instead of verbal categories can be used with the system. The pictures are designed to help children who have not yet acquired reading skills complete the task. Each child received a detailed explanation of the procedure before starting measurements. The individual images on the scale were explained and linked to a loudness category (e.g., soft, loud, etc.). The children were told that, if the sound they heard fell between two main categories (e.g., very soft and soft), they should press the button between the two images (see Fig. 2). In order to familiarize the child with the stimuli, a test run with five examples were completed. The children were also told to judge the loudness of each sound individually and not to compare across presentations. Testing was carried out monaurally on the right ear of the normal-hearing group and the better hearing ear of the hearing-impaired group. Although in typical situations the system automatically selects the test frequencies based on audiometric configuration, for this trial 0.5 kHz, 1 kHz, 2 kHz and 4 kHz were manually selected. This was to allow sufficient measuring points to judge the reliability and validity of the procedure with children. The LPP procedure provides three levels of accuracy: standard, extended and precise. These terms, which are selectable via the software, relate to the number of stimulus presentations at each frequency (i.e., standard has the least presentations and precise the most). Testing was conducted both in the standard and extended modes in the trial. All children carried out the procedure twice. To prevent the children from being distracted, care was taken to ensure they were not able to see the computer monitor. They were encouraged to make use of the repeat button if they were unsure of the loudness of a sound. The start/stop button allows a temporary interruption of the procedure. This was used by the experimenter to actively intervene during the procedure if any additional encouragement or explanation was required. Results and Discussion
The standard mode has been judged by the authors to be a reliable procedure for collecting data from adults. However, the collected data points using this setting were not sufficient to provide similar accuracy for children. Selecting a mode with more presentations (i.e., the extended mode) improved the results significantly and led to greater accuracy. The extended mode requires approximately 10 minutes per ear for three frequencies. We observed that the children were generally able to maintain good concentration for approximately 10 minutes. When given a short break with refreshment between the testing of each ear, it was possible to complete the procedure without difficulty.
Special attention was paid to the question of the use of the entire scale, as we were concerned that children may cluster their responses at one end of the scale only. The results, therefore, were analyzed relative to how often each category was selected. The study shows that children used the scale in a very similar manner to adults. Not only did they use the entire scale appropriately, they also used main categories more than the sub-categories in a manner comparable to the average adult data (Fig. 5). This suggests that the children in our sample understood the task. The allocation of an individual button to a loudness category, together with intermediate categories, appeared to increase the childrens confidence. Also, the repeat button on the ARU allowed the children to bridge short lapses in concentration and further increased their confidence. One feature of the procedure that we found particularly useful when testing children was the delay between presentations; the system waits for a response before presenting the next stimulus, therefore adapting to the timing of individual subjects. Each child completed the procedure twice, which allowed evaluation of test/re-test reliability. The childrens results showed good reliability and proved that the procedure is valid for use with children at least seven years old. Further study should evaluate whether younger children are able to complete the task. Generally, the children in the trial tended to be more spontaneous in their approach to the procedure than adults and were less concerned with the next signal. This meant they were often more capable of judging the loudness of each sound separately (thus avoiding the comparison of signals). It is, however, important that time is taken to clearly explain the procedure, as this will influence the accuracy of the results. Summary In this field study, the Loudness Perception Profile (LPP) was shown to offer a useful additional tool when programming hearing instruments for children over the age of seven years. The results obtained in this trial were similar to those found in adults and, on average, children displayed good test/re-test reliability. In most cases, the audiogram and the selection of a validated fitting procedure, such as the Desired Sensation Level (DSL [i/o]), is all that is required to successfully program Claro hearing instruments for children. LPP was found to be particularly useful as an additional tool in cases in which children are exceptionally sensitive to loud sounds. This article was submitted to HR by Andrea Bohnert, MTA, and Petra Brantzen, MTA, who are audiologists affiliated with the Clinic for Communication Disorders at Mainz University Hospital in Germany. Correspondence can be addressed to HR or to: Andrea Bohnert, MTA, Mainz Univ. Hospital, Clinic for Communication Disorders, Langenbeckstrasse 1, Building 403, 55101 Mainz, Germany; email: bohnert@kommunikation.klinik.uni-mainz.de. References |
|
|
|