Issue StoriesComplaints About Hearing in Noise: A New Answerby Laurel A. Christensen, PhD; Dana Helmink, MA; Wim Soede, PhD; and Mead C. Killion, PhD Much has been written about the benefits of directional microphone hearing instruments and how to fit these devices. However, directional instruments are still underutilized. This article details some of the reasons for this, and describes a new type of directional device: an array microphone that uses a spatial distribution of three microphones to achieve its directivity. Candidacy and operational issues are addressed, and two field studies are used. While spending the weekend with my father, I had the opportunity to see a friend of my parents whom I hadnt seen in 30 years. I remembered her from my childhood, and I really looked forward to seeing her again. Unfortunately, she missed most of what I said. At one point in the conversation, she mentioned that she had tried a hearing aid 12 years ago, but did not keep it because it worked so poorly in background noise. I tried to tell her about improvements in the field, but she explained that her friends with new hearing aids had the same complaints.
I felt sad knowing that a hearing aid with a directional microphone or an assistive listening device, such as an array microphone or FM system, could completely change her life. Unfortunately, she is part of the nearly 80% of people who report hearing difficulty but do not use hearing instruments.1 She seems likely to stay in that group, despite my urging her to give hearing aids another try. Additionally, I received a call from my sister-in-laws father. As he had done several times in the past, he inquired about the latest in hearing aid technology. (He has worn programmable analog hearing aids for the past 10 years.) At the time he purchased these aids, they were considered state-of-the-art. Unfortunately, he has never been truly satisfied with them, even though he wears them most of the time. His biggest complaint is his inability to hear in noisy situations, and he avoids noisy places and groups of people because he cannot communicate in such an environment. These two experiences, related by author, Laurel A. Christensen, PhD, share one fact: both individuals complained about their inability to hear in noise, one of the most common complaints of hearing aid users.2 Fortunately, in the past 5 years, solutions to this problem have been developed. The better directional microphone hearing aids can return many users to normal ability to understand speech in noise.3 Unfortunately, these solutions are underutilized by dispensing professionals. In fact, only an estimated 10% of hearing aids dispensed in the US include directional microphones (estimate from industry data). Much has been written about the benefits of directional microphone hearing aids and how to fit these devices. Then why are directional microphones so seldom ordered? We offer the following possibilities: 1) Until recently, hearing aids with directional microphones were only available in BTE and full-shell ITE styles, and few dispensing professionals realize that a 5 dB signal-to-noise ratio (SNR) improvement is now available in mini-canal sizes; 2) Most dispensing professionals do not know how to evaluate a patients SNR loss (the reduced ability to hear in noise), even though excellent tests are available;4,5 3) Some hearing aids with directional microphones are not very directional. We have measured 14 ITE hearing aids with directional microphones. Their directivity, or SNR improvement, varied from 2.2 to 5.1 dB; 4) The interaction between venting, low frequency response, and directivity is not well understood. Some hearing aids roll off the lows in the directional mode. Without sufficient low-frequency gain to overcome the noise entering through the vent, patients with low-frequency loss: a) become deprived of low-frequency cues; and b) lose the directional advantage at low frequencies;6 5) Until now there has been no practical head-worn device for the estimated 35% of patients whose SNR loss falls between 5 dB and the 15 dB that requires an FM system. The Link-it array microphone system is designed to provide approximately twice the directivity of a well-designed single-cartridge or dual-omni-directional microphone hearing aid. The development of the device began in 1986, as part of co-author Wim Soedes PhD thesis at the Delft University of Technology, The Netherlands.7 This technology was then developed into a practical wearable device by Etymotic Research and ARDEA during a collaboration that began in 1995. The Link-it is worn over the ear and requires no external cables or hand-held devices. It can be used with either ITE or BTE hearing instruments, requiring only that the hearing aids have built-in t-coils (Figure 1). A vertical orientation of the t-coil works best for an ITE hearing aid. Using two Link-its will provide binaural processing, which will typically provide 2-3 dB additional advantage over those devices that feed the same mono signal to both ears. The Link-it will also work with a cochlear implant through direct audio input (DAI), and can be used like a CROS system.
What is an Array Microphone?
An array microphone is a highly directional device that uses a spatial distribution of three or more microphones to achieve directivity. The greater the spacing, the better the directivity. Figure 2 shows three directional microphones spaced approximately 25 mm apart from each other, a distance that is much greater than normally used in hearing aids, but short enough to make a head-worn array practical.
Directivity is achieved using the three directional microphones. Figure 3a shows that a sound wave arriving from the front will have different arrival times at the three microphones. In the Link-it, each microphones output signal is delayed relative to the next one by approximately 75 microseconds, the amount of time it takes sound arriving from the front to travel between microphones with the spacing shown. As a result, all three waveforms will be in phase when added together. The resulting signal is three times as large as the original signal.
In contrast, in Figure 3b, a sound wave is shown arriving from the side. For this sound, the wave arrives simultaneously at each microphone, but each microphone still applies a delay to the signal. This results in the output signal from the three microphones being out of phase so they partially cancel each other out. This delay-and-sum methodapplied to three microphone elements that are already directionalcan provide an effective noise reduction of 7 dB over an ITE omni-directional microphone, and 8-10 dB over a BTE omni-directional microphone. More precisely, when measured on the KEMAR, the Link-it gives an average AI-DI of 7 dB. BTE omni-directional microphones typically measure -1 dB to -3 dB. The directivity vs frequency of the system can be seen in Figure 4. Candidacy for Array Mic The first step in determining candidacy for a Link-it is to evaluate the patients ability to hear in noise, or to determine a patients SNR loss. This can be done using the Quick SIN test or the HINT test.4,5
A measure of SNR loss is a simple procedure and should be a standard part of every hearing evaluation. Figure 5 shows previously published data of hearing loss as a function of the ability to hear in noise (signal-to-noise ratio for 50% correct identification of words-in-sentences). This data demonstrates: 1) SNR loss is not well predicted by hearing loss (ie, audiogram), because patients with mild hearing loss may have severe SNR loss and vice versa; 2) Each 1 dB of SNR loss creates a 5%-10% loss of intelligibility, depending on SNR loss and listening conditions. As seen in Figure 5, SNR for 50% correct generally increases as hearing loss increases; however, it cannot be predicted entirely from hearing loss. Looking at the point on the graph for a pure-tone average of 60 dB HL (marked by second arrow), the SNR testing for four patients can be seen. Results for these patients are 4.5 dB, 9 dB, 12 dB, and 15 dB. All have the same average hearing loss, but very different abilities to hear in noise. A normal hearing listener will score 2 dB on this test. Thus, the SNR loss for these four patients is 2.5 dB, 7 dB, 10 dB, and 13 dB, respectively. The solutions for hearing in noise are well represented in Figure 5. The patient with an SNR loss of 2.5 dB can be brought back to normal (possibly better than normal) hearing in noise with a directional microphone hearing aid. The two patients with SNR losses of 7 dB and 10 dB will probably also appreciate a hearing aid with directional microphones, but will still have trouble hearing in noise with the hearing aid alone. These two patients would be ideal candidates for an array microphone, to be used situationally when their directional microphone hearing aid is not enough. The patient with a 13 dB SNR loss would also appreciate an array microphone, but would benefit most from an FM system. Operation of Device The Link-it operates on a size 10 battery, which gives 1 month of battery life when used 4 hours per day. In use, the device is placed over the ear, switched on, and the hearing aid is switched to the t-coil position. The volume control on the hearing aid can be used to adjust the volume on the Link-it, and there is a screwdriver-set gain control on the device. The array microphone system was designed with a high magnetic field output to overcome electromagnetic interference from fluorescent lights and computer monitor screens. The 1000-2000 mA/m minimum field strength of the Link-it is more than 30 times the typical 30 mA/m field strength of hearing-aid-compatible telephonesenough to overcome most interference. If interference does become a problem, the user can temporarily switch in an additional 8 dB (2500-5000 mA/m) of field strength. Conclusion A head-worn array microphone can provide dramatic benefit with relatively easy operation for those patients whose SNR loss exceeds the reach of high performance directional microphone hearing aids. SNR losses between 5 dB and 15 dB may be well served by array microphones. The important first step is to determine SNR loss in order to properly advise patients about their options.
References 1. Kochkin S. The VA and direct mail sales spark growth in hearing aid market. Hearing Review. 2001;8(12):16-65. |
|
|