Issue StoriesIs the End Near for Acoustic Feedback?by Chester Pirzanski, BSc, and Brenda Berge, AuD Digital shell-making technology holds great possibilities for the fitting process. However, this technology will be limited by the same factors that limit traditional impression-taking technologies: a need for the use of open-mouth impressions and higher viscosity impression materials. Digital shell manufacturing and digital imaging of the ear canal hold excellent promise for improving the fitting process. However, these technologies will be limited by the same factors that currently limit the success of the more traditional impression-taking technologies: asymmetrical ears, the dynamic ear canal, and current impression-taking techniques. This study shows that a patients ears are symmetrical relative to being soft or firm (mobile or fixed) only 38% of the time. About 20% of all ears are mobile or soft (which also roughly corresponds to the the overall ITE return rate), but its very unlikely that a dispensing professional can recognize a soft/mobile ear through casual observation. Employing the open-mouth ear impression technique and using higher-viscosity impression material would radically reduce return rates. Feedback is one of the most difficult problems to overcome when fitting a person with a severe or profound hearing loss. This problem is not new and has been with us since the commercial introduction of portable, electronic hearing aids in the 1930s. Audible oscillation in a hearing instrument, also known as acoustic feedback, occurs when a portion of the output signal from the receiver radiates back to the microphone. If the radiation occurs through a vent, the feedback may be controlled by plugging the vent or reducing its size. However, in most cases, the radiation occurs around the earmold, which provides unintentional, uncontrolled, and inconsistent venting conditions. Earmold labs constantly look for new materials and technologies that would increase the accuracy of the earpiece fit in the patients ear and prevent the occurrence of acoustic feedback. This research is not without its challenges. The Acrylic Mold The Vinyl Mold To improve the situation, earmold labs attempted to use modified vinyl that expanded under body temperature and provided a tight seal in the ear canal within minutes. While some patients benefited from this invention, many reported excessive tightness, and the project was terminated. Obviously, this did not stop research for materials that would provide both adequate seal and comfort. The Silicone Mold Practical clinical experience with the fittings quickly revealed that silicone molds were unable to eradicate acoustic feedback, loose fit, and comfort issues from the list of failures due to earmolds. While many dispensing professionals regard silicone earmolds as indispensable, many others use hard acrylic molds just as successfully. A study4 investigating whether dispensing professionals fit soft or hard earmolds more successfully found that soft earmolds required 0.6% more remakes than hard earmoldsa finding that was unexpected. Some consolation can be found in the fact that 39% of the soft earmolds in this study were ordered for profound and severe hearing losses, whereas fewer (24%) hard earmolds were ordered for such fittings. Research designed to establish the underlying causes for poorly fitting earmolds found that even the most experienced technicians could fail to make a properly sealing earmold from an excellent-looking impression, but might also unexpectedly succeed with a poor impression. It was also found that, if several technicians processed replicas of an ear impression, the resulting earmolds looked somewhat different. Typically, the earmolds varied in the length of canal, the extent of impression trimming, and the thickness of wax coating. Inconsistencies in earmold manufacturing surfaced as the likely reason for poorly fitting molds. The Digital Mold In this new digital shell-making process, a laser scans the impression taken by a dispensing professional. Next, the digital image of the impression appears on a computer screen where it is modeled and altered digitally using specialized software into the final hearing aid shell or an earmold. Finally, a printer builds the earpiece from a pool of vinyl powder or light-curing liquid resin.5 These processes of impression duplicating, shell modeling, and printing are done with extreme accuracy. (For a review of the digital shell-making process, see Cortez et al.6) The first digital shells for hearing aids were built as exact replicas of the impressions. Impression trimming and waxing associated with the traditional technology (and commonly blamed for altering the finished mold shape) were expected to disappear forever. Unfortunately, the clinical fitting results were a surprise to many: the digital earpieces were often difficult to insert into the patients ear and uncomfortable. To remedy comfort issues, impression trimming was restored and...acoustic feedback surfaced as a major problem. To combat this, hearing aid shells were made tighter by enlarging the virtual ear impression with a surface offsetwhich is the computer-age equivalent of good-old waxing. Today, the results of digital shell-making technology are still not as good as was first expected. Seeing these mixed results, some hearing aid companies are delaying making digital molds for their customers. It should be emphasized that this isnt related the technology behind manufacturing digital shells (which is good); rather they fear that poor ear impressions taken by dispensing professionals compromise the fitting of digital shells. Projects have been undertaken to design a device that would take a direct scan of the patients ear canal and email the complete and undistorted image of the ear directly to the earmold lab. This image would then be processed with specialized software. In this approach, there would be no impressions with short canals, voids, wrinkles, etc. The idea is that the perfect ear image would yield a perfectly fitting earpiece. This, many contend, would be the end of acoustic feedback. Digital Imaging Would End Acoustic Feedback...Or Would It? However, in the authors opinion, the potential for reducing hearing aid remakes is greater than commonly thought. In fact, the hearing care field already has all the necessary materials, techniques, and technologies to eliminate most of the rework.8-10 We only have to use these resources more effectively and consistently. Digital shell-making technology and digital ear imaging have great potential. However, even digital ear imaging has its limitations in reducing return ratesjust as our current technology has its limitationsif the dynamics of the ear canal are not considered. Why Digital Imaging Benefits Would Be Irrelevant Unless We Account for the Dynamic Ear Canal
Evidence points to the potential to lower the volume of hearing aid remakes in the first quarter from 55% to 20%or even 15%. The impact of such a reduction on customer satisfaction would be enormous. Additionally, the manufacturing cost of hearing aids would be significantly reduced with a reduction in return rates, and these savings could be passed along to consumers in the form of lower hearing aid prices. To accomplish this, the root causes for fitting failures at the beginning of hearing aid service need to be established. Analyzing Binaural Fittings. Recent studies on remakes occurring for binaural hearing aid fittings provide an interesting insight into the nature of the remakes. Binaural earmolds are interesting to analyze because they are made from ear impressions taken at the same time, from the same patient, with the same material/technique, and by the same clinician. At the lab, the same technician will use these impressions to make shells, and then case and finish the aids. This simultaneous processing eliminates many variations that occur in monaural fittings where impressions are taken from different subjects, on different occasions, and the aids are manufactured by different labs and technicians. Based on this parallel processing, one might assume that, if one aid in a binaural fitting requires a remake, the other would also require a remakemost likely for the same reason. In other words, in binaural fittings, one would expect that binaural remakes should be the most common. Using data from practicing audiologists enrolled in the PCO School of Audiology AuD program, the authors investigated 337 binaural fittings where one or both instruments required a remake. The patients had similar hearing losses in both ears, and the aids were built with consecutive serial numbers. The results showed that the returns were almost equally divided between binaural, right ear, and left ear remakes. In fact, binaural remakes were the least common, accounting for only for 32% of the returns. Remakes for left ears accounted for 35% of the returns; and remakes for the right ears accounted for 33%. Surprising Patterns in Digital Shell-Making Returns. A similar pattern in remakes also emerged unexpectedly in digital shell manufacturing: The number of binaural hearing aid fittings is not reflected by the number of binaural remakes. While binaural orders are common for new aids, remakes are typically for only one aid. Obviously, this is not because binaural aids fit better; it is because a remake is usually required for only one aid in the pair. Figure 2 shows a digital modeling of two hearing aids done with 3Shape software.5 Great care was taken in the shell design. Soon after fitting, one aid was returned due to feedback problems. How could that be? The ears appeared symmetrical, both impressions were satisfactory, and the aids were modeled like twins!
The reason for the remake was that the impressions were taken with a soft silicone while the patients mouth was relaxed (closed). This impression technique was inappropriate for this particular patient. In this case, his right ear was firm, whereas his left ear was softer and had an active ear canal. Not surprisingly, the left aid failed relative to acoustic feedback. Explanation: Human Ears Are Asymmetrical It was found that, for some subjects, one ear was firm and not active while the other ear was soft and severely affected by mandibular movement. In fact, the variances in ear tissue softness and mandibular movement between the two ears of one subject were as great as variances found among ears of different subjects. These results are consistent with data obtained in previous studies.11,12 The fact that over 60% of patients have asymmetrical ears corresponds surprisingly well with the fact that over 60% of remakes occurring in binaural fittings are for one aid only. Since a loose fit and feedback are the dominant problems in hearing aid remakes, it is reasonable to conclude that the majority of ear impressions taken by dispensing professionals do not adequately capture the ear softness and the magnitude of mandibular movement. Another pair of factors that match is that approximately 80% of human ears are firm and not affected by mandibular movements,9 and that 80% of hearing aids do not require a remake.7 This implies that the 20% of ear canals that are soft and/or mobile should be recognized as such prior to impression taking. To lower the risk of fitting problems, impressions from such ears should be made with a firmer (higher) viscosity silicone and with the patients mouth wide open (see sidebar).13-14 This is not required for the remaining 80% of ears because impressions from firm and inactive ear canals can be taken with any impression material and any technique, and the resulting hearing aids will (almost) always fit well. The problem is that dispensing professionals are only able to successfully detect high-risk ears with a relatively low (25% to 46%) accuracy rate.9 In fact, many of the experienced audiologists evaluating the effects of altering the impression technique were surprised that they were unable to predict the changes in impression size and shape when examining patients ears.6 The Only Way to Go. This lack of precise assessment means that all ear impressions can and should be taken as higher-viscosity open-mouth molds. The common employment of this approach has the potential to eliminate more than a half of remakes in new hearing aids and significantly elevate patients satisfaction with their instruments.15-17
Summary In this light, it can be seen that digital shell-making technologywhile extremely helpful and efficientwill also suffer from many of the same limitations that our traditional shell-making technologies have relative to the elimination of acoustic feedback. Whatever the technology, for the best fitting results, clinicians should modify their impression technique and routinely take open-mouth impressions with a firmer silicone for all hearing instruments. When this is accomplished, the end of acoustic feedback will be much nearer.
Correspondence can be addressed Chester Pirzanski, Starkey Labs Canada, 7310 Rapistan Ct., Mississauga, Ontario, L5N 6L8, Canada; email: cpirzanski@hotmail.com. References |
|
|
|
ADDITIONAL ONLINE RESOURCES |
|