Issue StoriesProving Adaptive Directional Technology Works: A Review of Studiesby Thomas A. Powers, PhD, and Volkmar Hamacher, PhD A review of and commentary on four recent important studies on the subject of adaptive directionality and dual- and three-microphone hearing instruments. Directional microphones are designed to collect sounds from a desired direction and attenuate sounds coming from other directions. They have been applied to hearing instruments for several years and have proven to significantly increase speech intelligibility in various noisy environments (see articles by Ricketts and Mueller1,2 for a review). As described by Powers & Hamacher,3 we have recently observed three major advances in directional hearing instruments: 1) the use of a three-microphone array; 2) automatic switching or transition between directional and omnidirectional modes, and 3) the use of an adaptive directivity polar pattern. This article will primarily focus on adaptive directivity, describing how this feature is designed, and reviewing four behavioral studies that have been conducted using subjects fitted with the Siemens TRIANO S (dual-microphone) and TRIANO 3 (three-microphone) hearing instruments. Design of Adaptive Directivity
As shown in Figure 1, the signal of the rear microphone is delayed and subtracted from the signal that is received by the front microphone. The directivity pattern of the system is defined by the ratio of the internal delay (Ti) and the external delay (Te) corresponding to the microphone spacing (typically 7-16 mm). In this example, the ratio was set to 0.57 resulting in a super-cardioid pattern, as shown on the far right of Figure 1. To compensate for the first-order high-pass characteristic introduced by the differential processing, an appropriate low-pass filter (LPF) is added to the system.
Figure 2 shows that the direction of lowest sensitivity (or directivity notch) can be controlled by the ratio Ti/Te. If Ti/Te is changed from 0 to 1, the spatial notch moves continuously from 90° to 180°.
This effect can be exploited to enhance the directional performance in various everyday situations with noise originating from a specific location. The adaptive directional microphone continuously adapts its directivity pattern by Ti/Te variation to the respective noise field so that the directivity notch matches the direction of noise arrival. In this way, the noise suppression effect is designed to be automatically maximized in a variety of everyday listening situations which often differ in direction of noise incidence (Figure 3).
To be accessible for the adaptation system, the internal delay has to be applied in the digital processing domain of the hearing instrument. However, this would require the realization of fractional delay filters. Since these filters would cause a high computational effort, the adaptive directional microphone of TRIANO hearing instruments uses an efficient method as proposed by Elko & Pong.4 As shown in Figure 4, the shape of the directivity pattern is managed by a weighted sum of the output signals of a bi-directional and a cardioid pattern. In the digital domain, these can be calculated simultaneously. The position of the directivity notch is directly (monotonically) related to the weighting factor. To be beneficial in the dynamic noise fields of everyday situations, the management of the directional notch has to be reliable and accurate. Additionally, it should not introduce artifacts or perceivable changes in the frequency response for the 0° (directly in front) target direction, which could be irritating or annoying for the user. Finally, the adaptation process must be fast enough to compensate for head movements and to track moving sources in common listening situations, such as conversation in a street cafe with interfering traffic noise. The algorithm implemented in the TRIANO adaptive directional system is designed to fulfill these requirements. The adaptation time for a noise source rapidly moving from 180° to 90° is less than 50 ms, which is sufficient for the dynamics of most listening situations. To ensure that no target sources from the front hemisphere are suppressed, the directivity notches are limited to the back hemisphere (90°-270°). Additionally, the depth of the notches is limited to prevent hazardous situations for the user (eg, when someone is crossing the street and a car approaches).
Figure 5 shows a measurement in an anechoic test chamber with an adaptive directional microphone hearing instrument (TRIANO 3 BTE, linear amplification mode, noise reduction deactivated) mounted on the left KEMAR ear. A noise source (500 Hz) was moved around the head and the output level of the hearing aid was recorded (blue line). Comparing the same measurement for a non-adaptive super-cardioid directional microphone (red line), one can clearly see the higher suppression effect for noise incidence from the rear hemisphere. While the laboratory data with adaptive directional microphone hearing instruments has been encouraging, it is also important to consider how these instruments affect the understanding of speech with human subjects. It is well known that electroacoustic benefits observed in an anechoic chamber might not translate to improved speech understanding in noise in real-world settings. In the past year, several studies have been conducted with the adaptive directional hearing instruments discussed, and four studies have been selected here for review. Effects in Diffuse Noise Field The results were similar for both test environments, so we will summarize the general findings. When the hearing instruments were tested in the fixed-directional mode, the mean L50 for TRIANO S and competitive aid were now essentially the same, both significantly poorer (by 2.5 dB) than that of TRIANO 3, and it was also superior to the competitive directional aid (about 2.0 dB). Similar results were obtained when the hearing instruments were set in the adaptive-directional mode, except that the mean L50 for TRIANO S and competitive aid were now essentially the same, both significantly less (by 2.5 dB) than that of TRIANO 3. The fact that the adaptive setting of TRIANO did not result in improved performance over the fixed setting was an expected finding: the noise was not moving, nor was it originating from a specific location. The finding showing that mean performance was no worse for adaptive-directional is a positive finding, as it shows that the adaptive algorithm was working correctly for a diffuse field environment. Noise From Various Discrete Azimuths
For testing in Room 1, there was increased benefit (SNR improvement) for the adaptive directional setting compared to fixed directional for both the dual microphone and three-microphone aids in nearly all test conditions (benefit around 1.5 dB to 2.0 dB SNR). For the dual-microphone instrument, there was a significant overall benefit for the adaptive setting, and individual significance for the 90°, 120°, 150°, and 180° azimuths (no benefit would be expected for the 0° azimuth). For the three-microphone instrument, the adaptive setting was significantly superior to the fixed setting for overall performance, and also for noise presented from 90° and 180° azimuths. As would be expected, because of the arrangement of Room 2 (16 surrounding loudspeakers), there was no significant difference between the fixed versus adaptive mean L50 scores for either of the directional instruments. Moving Competing Noise Source The results showed a significant advantage (HINT score improvement of approximately 2 dB) for the adaptive directional setting (compared to fixed directional) for the competing stationary noise condition for both models. To some extent, this finding was expected, as the 160° azimuth of the stationary noise does not correspond to the null of the polar plot in the fixed directional mode. More importantly, a significant improvement (HINT score change of approximately 2 dB) was also observed for the adaptive directional processing for the moving noise condition. Across the two instruments, mean HINT scores for the fixed directional setting were about 5.7 dB better than omnidirectional, and scores for the adaptive directional setting were about 7.6 dB better than omnidirectional. Comparison to Normal-Hearing Individuals The authors used a modified version of the HINT, and testing was conducted using both stationary and moving-noise competing signals. Speech was presented at 0° azimuth. Six loudspeakers, placed at the the top and bottom corners of the test room (90°, 180° and 270° azimuths), were used to present the uncorrelated competing noise from the HINT CD (70 dBA). Noise was presented simultaneously from all loudspeakers for the stationary noise condition. For the moving noise, only 1 speaker at a time presented 2 seconds of noise; there were no silent intervals and the loudspeaker presentation was randomized. Testing was conducted for both a group of unaided, young, normal-hearing subjects (n = 48; mean age = 22) and a group of older hearing-impaired individuals (n= 46; mean age = 62) fitted with TRIANO S and TRIANO 3 instruments. The adaptive directional condition was not tested in the stationary noise environment (ie, no benefit would be expected). However, we will briefly report these findings so that they can be used as a reference for the moving-noise results. In the stationary noise condition, the hearing-impaired subjects wearing the test instruments did not perform differently than the young, normal-hearing subjects. On an individual basis, all but 4%-6% of the subjects performed within the range of the normal-hearing subjects (within ±2 standard deviations).
The results for the moving-noise testing are shown in Figure 6. As shown, there was a slight improvement in mean HINT scores for the adaptive versus fixed directional setting for both test instruments, but this difference was not significant. The adaptive setting of the TRIANO 3, however, provided enough directional hearing improvement to raise the hearing-impaired groups mean HINT performance to a level so that it was not significantly different from that of the normal-hearing (younger) subjects.
It is also important to consider individual data. Figure 7 shows the percent of subjects whose aided HINT score fell within the range of the normal-hearing subjects (within ±2 standard deviations) for the moving-noise condition. Observe that, for both the dual- and three-microphone instruments, there was a slight improvement for the adaptive setting (87% vs 91% and 89% vs 94% respectively). Moreover, these results show that, when adaptive directional technology is used, over 90% of older hearing-impaired adults perform within the range of normal-hearing individuals when a moving noise is present. Summary