Issue StoriesRECD and Clinical Verification in Childrenby Andrea Bohnert, MTA, and Petra Brantzen, MTA A field study on the accuracy of a new system for making direct real-ear-to-coupler difference (RECD) measurements in children. The real-ear-to-coupler difference (RECD) is the difference between the 2cc-coupler values and the individual auditory canal values measured with the identical sound pressure and conditions. RECD values are given in dB and for every frequency. Individual adult RECDs often conform closely to the average. However, childreneven of similar agefrequently have very different RECDs.1 Therefore, RECD measurement is vital in achieving the best possible hearing instrument fitting, especially in children. Moodie et al.,2 described a procedure currently used for RECD measurements with children. Research in children showed significant differences in sound pressure levels with and without individual RECD corrections.3 Until now, RECD measurements have become a standard tool in pediatric fittings. A number of companies have implemented an age-related average RECD in their fitting software. The pre-calculation for target (first fit) takes into account a smaller auditory canal volume as compared to adults. These average RECD values are used unless individually measured RECDs are entered. Phonak offers an optional fitting tool for the Supero high-power hearing instrument: RECDdirect. The fitting software provides an integrated measurement procedure which allows for direct RECD measurement, incorporating the users earmold. Except for the RECDdirect module, which is attached to the hearing instrument (Figure 1), no additional equipment is required. Based on this measurement, the gain and output of the hearing aid is calculated according to the patients individual auditory canal characteristics.
To clinically evaluate the RECDdirect fitting tool, a pediatric study was conducted which addressed the following questions:
Accuracy of RECD Measurement
RECDdirect was measured twice for each subject (in most cases, the measures were made by different audiologists). RECDs also were measured using a standard measurement system (Audioscan RM500) using the subjects individual earmolds in a subset of 35 subjects. Results: The reliability of RECDdirect measures (ie, the test-retest deviation) for the two age groups is shown in Figure 3. Average test/retest deviation for both groups was close to 0 dB. The standard deviations are lower for the older group, because the measurements were not influenced by movement or babbling, as in the younger children. In 5 of the very young children, the measurement was possible, but not reliable, because the children were too noisy. Two of these 5 children had multiple disabilities.
Comparison to Standard Test Equipment: In 35 children, the RECD also was measured with standard measurement equipment (Audioscan RM500), using the childrens own earmolds. The comparison between both systems is not straightforward due to the fact that the measured frequencies and their resolutions differ between both procedures. While the Audioscan uses warble tones with high frequency resolution, RECDdirect utilizes narrow-band noises with 5 different center frequencies. Thus, RECDdirect values at the 5 center frequencies were compared to the respective values of the Audioscan RECD curve. The average differences between RECDs measured with RECDdirect and Audioscan are shown in Figure 4. In general, the differences between both methods are small (average difference across all frequencies is 1.3 dB). Deviations in the low frequencies are largely due to leakage around the earmold.
Comparison to Age-Related Average RECDs: When individual RECDs are not measured, the fitting software assumes average values, which depend on age and venting. These average values do not consider individual characteristics of the ear canal. Figure 5 shows the differences between individual RECDdirect measures and the respective age-and vent-related average values. The standard deviation of the difference across frequencies was between 3 dB and 5 dB. In some cases, the childs individual RECD differed considerably from average values. This emphasizes that it is, in fact, a good idea to use individual RECDs in pediatric fittings.
Practical Considerations The fitting software automatically finishes the measurement as soon as a certain level of confidence is achieved. If a child is too restless, the measurement will be aborted after awhile and cannot be used for the fitting. This was the case in 5 children. On average, these children were 2 years old. Two of them had additional multiple disabilities. A solution to this problem could be a software button for the examiner to manually control the duration of the measurement for better use of the childs silent moments.
Summary The data from this field study also support the routine measurement of individual RECD values in children due to the variable size and shape of childrens ear canals, which often differ considerably from age-related averages.