Perform Real-ear Testing on Open-fit Hearing Instruments in two Steps

The elimination of the occlusion effect has encouraged most hearing instrument manufacturers to build open-fitted products. As will soon be pointed out in a paper titled, "Real-ear measurement verification for open, non-occluding hearing instruments," written by Johnannes Lantz et al and soon to be published in the International Journal of Audiology, the use of the pressure method of soundfield control can yield incorrect results when testing open fittings. There is hope, however. The method below shows a way that the verification can easily be accomplished with FONIX analyzers.

Two Easy Steps for Real Ear Testing on Open-Fit Devices
Use the normal recommended setup for the real ear test. Place the reference and probe microphones on the head as usual, and locate the soundfield speaker as recommended. Level the sound field with the ear unaided and, if desired, make an unaided response measurement (REUR). Now take the following steps:

Step 1. Turn off the reference microphone.

Step 2. Place the hearing aid on the ear and take the real ear aided response (REAR).

That’s it.
It is necessary to instruct the patient to hold still when making the REAR measurement with the reference microphone off. The hearing aid must be switched off while leveling the sound field.

The FONIX 7000 graph shows three curves. Curve 1 is of the REUR; Curve 2 is REAR with the reference microphone left on;  Curve 3 is REAR with the reference microphone off. Signal levels were all 65 dB SPL RMS, digital speech, ANSI weighting. The additional gain at 2.5 kHz with the reference microphone left on is the result of measurement artifact.

For information, contact:
Frye ElectronicsTigard, Ore, (800) 547-8209 www.frye.com

|
|

MEDIA CENTER

Interactive Media
Resources
Classifieds
Calendar
Consumer Resources
Media Kit
Advertiser Index
EAB
Reprints
Submit an Article

ADDITIONAL ONLINE RESOURCES

Allied Healthcare
Medical Education
24X7mag
Chiropractic Products Magazine
Clinical Lab Products (CLP)
Orthodontic Products
The Hearing Industry Resource
HME Today
Rehab Management
Physical Therapy Products
Plastic Surgery Products
Imaging Economics
Medical Imaging
RT Magazine
Sleep Review
SynerMed Communications
IMED Communications
Practice Growth
Practice Builders
powered by:
Copyright © 2008 Ascend Media LLC | The Hearing Industry Resource | All Rights Reserved.
Privacy Policy | Terms of Service