Issue StoriesSPECIAL "HOW TO" SECTION
Perform Real-ear Testing on Open-fit Hearing Instruments in two Steps Two Easy Steps for Real Ear Testing on Open-Fit Devices Step 1. Turn off the reference microphone. Step 2. Place the hearing aid on the ear and take the real ear aided response (REAR). The FONIX 7000 graph shows three curves. Curve 1 is of the REUR; Curve 2 is REAR with the reference microphone left on; Curve 3 is REAR with the reference microphone off. Signal levels were all 65 dB SPL RMS, digital speech, ANSI weighting. The additional gain at 2.5 kHz with the reference microphone left on is the result of measurement artifact. For information, contact: |
|
|